Related Products: |
EGSK-201, EGMA-203 |
Related Documents: |
EZDP-2046 |
Type: |
Maintenance |
Keywords/Labels: |
Remote Meter Point, Calibration, Maintenance |
Summary: |
Calibration procedure for the gain trimpot. |
The gain trimpot is factory set. Do not adjust it unless specifically instructed to do so by Cutsforth Support.
To adjust the gain trimpot
Turn the trimpot
in any direction to the farthest extent until there is an audible click.
Turn the trimpot in the other direction for 15 turns and mark the position with a permanent marker or similar.
![trimpot-gain-412px.png](/hc/article_attachments/360058028292/15ecd3d23f107f.png)