Related Products: |
EGMA-004, EGMA-005, EGMA-006, EGMA-007, EGMA-008, EGMA-013, EGMA-015, EGMA-016, EGMA-019 |
Related Documents: |
EZDP-2020 and EZDP-2038 |
Type: |
Technical information |
Keywords/Labels: |
Premium Monitoring System, Calibration, Set up, Maintenance |
Summary: |
Advised calibration frequency for Premium Monitoring System touch screen as preventative maintenance. |
Advised Frequency: Every three years or anytime it is observed through experience that the touch screen accuracy is not acceptable.
Action:
Exit the Premium Monitoring System Application.
On the Windows Desktop, in the lower-right corner of the screen next to the system date and time, find the icon for the PenMount Monitor.
Right-click the PenMount Monitor icon and click Control Panel.
Highlight the PenMount 6000 USB icon and click Configure.
Click Standard Calibration.
Follow the prompts to complete the calibration.
Download Snapshot Waveforms and Baseline Recordings
Calibrate Power Supply of 120 V to 24 V AC/DC Converter
Calibrate Wear Indicator Voltage Sensors or Ground Current Sensor
Check Current Sensor, Voltage Sensor, and System Performance